| Regulation, Specification, or Guidance
 | 
Requirement
 | 
| ASTM E1578-18 E-3-1
 | 
10.1 The system should provide a means for tracking usage of laboratory equipment and instruments.
 | 
| ASTM E1578-18 E-3-2
 | 
10.2 The system shall provide a means for planning the use of and reserving equipment and instruments.
 | 
| ASTM E1578-18 E-3-3
 | 
10.3 The system shall be able to bring an instrument online and take an instrument offline both manually and via a configurable scheduler.
 | 
| ASTM E1578-18 E-3-4
 | 
10.4 The system should record instrument usage time to assist laboratory personnel with capacity planning and scheduling.
 | 
| ASTM E1578-18 E-3-5
 | 
10.5 The system shall allow users to configure and record multiple instrument events for the same instrument, including different event types and different frequencies for the same event type.
 | 
| ASTM E1578-18 E-3-6
 | 
10.6 The system shall be able to group instruments together in specific ways, including by type and laboratory location.
 | 
| 
 21 CFR Part 211.67–68 
21 CFR Part 211.160 (b-4) 
21 CFR Part 211.30 (b) 
21 CFR Part 211.60 (e) 
21 CFR Part 225.30 (b-4) 
21 CFR Part 606.60 
21 CFR Part 820.70 (g) 
21 CFR Part 820.72 
42 CFR Part 493.1252 (b-3) 
42 CFR Part 493.1255 
AAFCO QA/QC Guidelines for Feed Laboratories Sec. 2 
AAVLD Requirements for an AVMDL Sec. 5.5 
ABFT Accreditation Manual Sec. E-20 
ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 5.6.1.1 
AIHA-LAP Policies 2018 2C.3.7–2C.3.10 
ASTM E1578-18 E-3-7 
EPA 815-R-05-004 Chap. IV, Sec. 7 
EPA ERLN Laboratory Requirements 4.11.11–13 
EPA QA/G-5 2.2.6–7 
ISO 15189:2012 5.3.1.4 
ISO/IEC 17025:2017 6.4.7 
ISO/IEC 17025:2017 6.4.8 
USDA Administrative Procedures for the PDP 5.2.4 
USDA Data and Instrumentation for PDP 5.2 
USDA Data and Instrumentation for PDP 6 
WHO Technical Report Series, #986, Annex 2, 13.5 
WHO Technical Report Series, #986, Annex 2, 16.23
 
 | 
10.7 The system shall allow the configuration of calibration and maintenance frequency and time frames, with available intervals of days, weeks, months, and years.
 | 
| 
 21 CFR Part 225.30 (b-4) 
AAVLD Requirements for an AVMDL Sec. 5.5.9 
ASTM E1578-18 E-3-8 
WHO Technical Report Series, #986, Annex 2, 16.23
 
 | 
10.8 The system shall provide clear alerts or notifications when an instrument nears its calibration due date.
 | 
| 
 21 CFR Part 211.160 (b-4) 
ABFT Accreditation Manual Sec. E-21 
ASTM E1578-18 E-3-9 
ISO 15189:2012 5.3.1.5 
ISO/IEC 17025:2017 6.4.9 
WHO Technical Report Series, #986, Annex 2, 13.10
 
 | 
10.9 The system shall clearly identify any instrument that is out-of-calibration, beyond its preventative maintenance due date, or under investigation and prevent it from being selected for use.
 | 
| 
 21 CFR Part 820.72 (b-2) 
ASTM E1578-18 E-3-10 
EPA ERLN Laboratory Requirements 4.11.11–13 
EPA QA/G-5 2.2.6–7 
ISO/IEC 17025:2017 6.4.8 
WHO Technical Report Series, #986, Annex 2, 16.23
 
 | 
10.10 The system shall be able to show all instances of scheduled calibration, preventative maintenance, and service dates for an instrument.
 |