LabLynx KB:Healthcloud POL user manual - data entry

From LIMSWiki
Jump to navigationJump to search

Data Entry Overview

Any Sample Groups/Batches (or Non-Batch Samples) in any process step are displayed on the Data Entry list screen. These are displayed in three ways via three tabs:


  1. Active Batches - This is the default tab displayed, showing all batches the user has profile access to view, by Process Group Number (Batch Number) showing which process the batch is for, created date/time, who created it and which (if any) instrument is associated with it.
  2. Active Processes - All batches the user has profile access to view, by Process showing Process Group Number (Batch Number), created date/time, who created it and which (if any) instrument is associated with it.
  3. Non Batch Samples - This is a list of samples whose assigned test includes the non-batching process (they are processed individually). The list displays sample number, specimen type, status and patient name.


Data Entry 1 875.png


Results may be entered manually by selecting the desired batch and then proceeding as described in this section. Results may also be entered automatically if instruments are interfaced with the LIMS.


To Enter Test Results

On the Data Entry list screen (main screen displayed by choosing Data Entry menu item), click on the batch that contains the samples for which you wish to enter data. If necessary, use the Filter to find it. The Batch Result Entry screen displays.


Data Entry 2.png


There are two ways to enter results data manually: (1) by Test and (2) by Sample.


To Enter Results by Test

On the Batch Results Entry screen, click on the Test for which you would like to enter results (this will display all Samples assigned that Test).


Data Entry 3.png


The Process Data Entry by Test screen displays. Enter data into Raw Results column for all desired samples.


Data Entry 4.png


Click Save Save Button.jpg. The LIS populates Calc and Report results and displays according to pre-defined sig figs and resolution.


Data Entry 5.png


  • NOTE: If any results are out of the pre-defined parameter range, the icon column will change color and the LIMIT EXCEEDANCE icon will display:


Out-of-range alerts

Data Entry 6.png


Hovering over the icon displays information about the exceedence. Additional information can be viewed by clicking the icon.


Data Entry 7.png


If any results have calculations associated with them, select them (click their checkbox) and click Calculate Calculate Button.png button to display calculated values.


Data Entry 8.png


Select all sample runs whose results are satisfactory and click Approve Approve Button.png. The selected samples are moved to be available for the next Process and no longer display.


To Enter Results by Sample

On the Batch Results Entry screen, click on the Sample for which you would like to enter results.


Data Entry by Sample 1.png


The Data Entry by Sample screen is displayed. This will display all Tests assigned to that Sample. *Note: Authorized users may add or delete tests here even at this stage if necessary.


Data Entry by Sample 2.png


Click the Data Entry By Sample Data Entry by Sample Button.png button. The Sample Data Entry screen is displayed.


Data Entry by Sample 3.png


Enter results in the Raw Result column for the tests for which you have results, and click Save Save Button.jpg.


If any results are outside of pre-defined parameters, a color-coded alert and icon will display. Hovering over the exceedence icon will display information about the exceeded limit.


Limit Violation

Data Entry by Sample 4.png


Hovering over icon will display more information about the limit.


Data Entry by Sample 5.png


Clicking the icon will display still more information about the limit.


Data Entry by Sample 6.png


To Complete Results Entry

Once data are entered and saved for all specimens and tests, click Commit Commit Button.png. The results no longer display and the specimens are complete for that batch.


Results Review

To Approve/Rerun/Cancel Results

Results may be approved, sent for re-test (another run created) or canceled (discard results). In the Batch Result Entry screen (accessed by clicking on a batch in the Data Entry list screen), click the Open All Open All Button New.png button.


All Tests and Samples are displayed with each parameter shown.

Data Entry Open All.png


Normally one or more Process steps of Review are created and added to a Path which is assigned to a given test and then assigned to appropriate specimens - and approval review is done in that process batch. However, it may also be done for any batch regardless of process step.


Results can be approved, sent for re-test (another run created) or canceled (discard results) either by individual sample for a given test (all analytes/parameters in that test), individual specimen for all tests including all analytes/parameters, or by individual analyte/parameter or for an entire test (all specimens in that batch).


To Approve, Send for Rerun or Cancel by Specimen/All Analytes, Single Test

To Approve, click ONCE in the yellow background area of the cell immediately below the Specimen Number where it says "No Action":

Data Entry Approve 1.png


The Results for that specimen for that test all turn green and the word "Approve" displays.


Data Entry Approve 2.png


Click the Commit Commit Button.png button to finalize.


To Send for Re-Test/Rerun, click TWICE in the yellow background area of the cell immediately below the Specimen Number where it says "No Action":

Data Entry Approve 1.png


The Results for that specimen for that test all turn yellow and the word "Re-run" displays.


Data Entry Approve 3.png


Click the Commit Commit Button.png button to finalize.


To Cancel Run, click THREE TIMES in the yellow background area of the cell immediately below the Specimen Number where it says "No Action":

Data Entry Approve 1.png


The Results for that specimen for that test all turn red and the word "Cancel" displays.


Data Entry Approve 4.png


Click the Commit Commit Button.png button to finalize


To Approve, Send for Rerun or Cancel by Specimen/All Analytes, All Tests in the Batch

To Approve, click ONCE in the yellow background area of the cell immediately below the Specimen Number where it says "No Action":

Data Entry Approve 1.png


The Results for that specimen for that test all turn green and the word "Approve" displays.


Data Entry Approve 2.png


Click the Commit Commit Button.png button to finalize.